Main parameters
Principle
WBC (IMG/Neu/Mon/Lym/Eos/Bas), NRBC/RET*, PLT-H/PLT-O*/IPF:
SF Cube ^ Cell Analysis Technology
^S: Scatter; F: Fluorescence; Cube: 3D analysis
RBC, PLT
- Flow-DC Impedance
HGB
- Colorimetric method
ESR
- Optical measurement method
Measured parameters (full sample): 109
Measured parameters: 41
WBC Bas# Bas% Neu# Neu% Eos# Eos% Lym# Lym% Mon# Mon% IMG# IMG% RET%* RET#* RHE* IRF* LFR* MFR* HFR* RBC HGB MCV MCH MCHC RDW-CV RDW-SD HCT NRBC# NRBC% PLT PLT-I PLT-H PLT-O* MPV PDW PCT P-LCR P-LCC IPF ESR
Research parameters: 68*
Measured parameters (body fluids)*: 18
Reporting parameters: 7
WBC-BF TC-BF BF# MN# MN% PMN# PMN% RBC-BF
Research parameters: 11
Sample volume
- CD (whole sample): 25ul
- CD-ESR (total sample): 160ul
- Pre-dilution: 20ul
Data storage capacity
- Up to 150,000 results including values and graphs*
Capacity
- CD 80h CDR 60h CD+ESR 40h
Analysis mode
Sample type |
Analysis mode |
Full form |
CBC, CBC+DIFF, CBC+RET, CBC+NRBC, CBC+RET+NRBC, CBC+ESR, CBC+RET+ESR, CBC+NRBC+ESR, CBC+RET+NRBC+ESR, CBC+IPF, CBC+RET+IPF, CBC+NRBC+IPF, CBC+RET+NRBC+IPF, CBC+ESR+IPF, CBC+RET+ESR+IPF, CBC+NRBC+ESR+IPF, CBC+RET+NRBC+ESR+IPF, CBC+PLT-O, CBC+RET+PLT-O, CBC+NRBC+PLT-O, CBC+RET+NRBC+PLT-O, CBC+ESR+PLT-O, CBC+RET+ESR+PLT-O, CBC+NRBC+ESR+PLT-O, CBC+RET+NRBC+ESR+PLT-O, CBC+IPF+PLT-O, CBC+RET+IPF+PLT-O, CBC+NRBC+IPF+PLT-O, CBC+RET+NRBC+IPF+PLT-O, CBC+ESR+IPF+PLT-O, CBC+RET+ESR+IPF+PLT-O, CBC+NRBC+ESR+IPF+PLT-O, CBC+RET+NRBC+ESR+IPF+PLT-O |
Diluted sample |
CBC+DIFF |
Body fluids |
BF, BF+DIFF |
Physical specifications
Size: 840D x 855W x 800H mm
Weight: ≤70.6kg
Electrical voltage: 100V-240V~ (±10%)
Frequency: 50Hz/60Hz (±1Hz)
Input power: 600VA
Connection port: LANx1, USB x 4 (Specifications: DC 5V; 500mA; USB2.0 x 3; USB3.0 x 1)
Normal operating environment
Temperature range: 10°C ~ 35°C
Relative humidity: 30% ~ 85%
Pressure
- 70.0kPa – 106.0kPa*
- *Note: required altitude for normal operation: -400m ~ +3000m
Efficiency
Parameter |
Linear range |
Precision |
Cross contamination |
WBC |
0-500×10⁹/L |
≤2.5% (≥4.51×10⁹/L) |
≤1.0% |
RBC |
0-8.60×10¹²/L |
≤1.5% (≥3.5×10¹²/L) |
≤1.0% |
HGB |
0-260g/L |
≤1.0% (110-180g/L) |
≤1.0% |
HCT |
0-75% |
≤1.5% (30%-50%) |
≤1.0% |
PLT* |
0-5000×10⁹/L |
≤1.5(SD) (≥20×10⁹/L)≤2.5% (≥100×10⁹/L) |
≤1.5% |
RET* |
0-8×10¹²/L |
≤15% (RBC ≥ 3.00×10¹²/L RET%: 1.00% ~ 4.00%) |
≤1.0% |
ESR |
≤1.8(SD) (0~20mm/h) |
≤1.0% |
Declaration: Applicable to CDR/PLT-O 5x and CR/PLT-O 5x series only
Items marked with (*) are only available on the BC-780.
Đánh giá
Chưa có đánh giá nào.