BC-700 & BC-720

Contact Us
Main parameters

Original reason

WBC (IMG/Neu/Mon/Lym/Eos/Bas), NRBC/RET*,PLT-H/PLT-O*/IPF: SF Cube ^ Cell Analysis Technology

^S: Scatter; F: Fluorescence; Cube: 3D analysis

RBC, PLT

Flow-DC Impedance

HGB

Colorimetric method

ESR

Optical measurement method

Measured parameters (whole blood): 109 Reported parameters: 41

WBC Bas# Bas% Neu# Neu% Eos# Eos% Lym# Lym% Mon# Mon% IMG# IMG% RET%* RET#* RHE* IRF* LFR* MFR* HFR* RBC HGB MCV MCH MCHC RDW-CV RDW-SD HCT NRBC# NRBC% PLT PLT-I PLT-H PLT-O* MPV PDW PCT P-LCR P-LCC IPF ESR

Research parameters: 68*

Measured parameters (body fluids): 18

Reporting parameters: 7

WBC-BF TC-BF# MN# MN% PMN# PMN% RBC-BF

Research parameters: 11

Sample volume

CD (whole blood): 25ul CD+ESR (whole blood): 160ul Pre-dilution: 20ul

Data storage capacity

Up to 150,000 results including values and graphs*

Capacity

CD 80t/h CDR 45t/h CD+ESR 40t/h

Analysis mode

Sample type

Analysis mode

 

Whole blood

CBC, CBC + DIFF, CBC + DIFF + RET*, CD + ESR, CDR + ESR*, CD/WBC-3X, CDR/PLT-5X*,

and other modes

Dilution money

CBC, CBC + DIFF, CDR*, and other modes

Body fluids

CBC + DIFF

Physical specifications

Size

500D x 325W x 450H mm

Weight

≤35Kg

Electricity pressure

100V-240V~ (±10%)

Frequency number

50Hz/60Hz (±1Hz)

Input power

300VA

Connection port

LAN x 1, USB x 4 (Specifications: DC 5V; 500mA; USB 2.0 x 3; USB 3.0 x 1)

Normal operating environment

Temperature range:

10℃ ~ 35℃

Relative humidity:

30% ~ 85%

Pressure:

70.0kPa ~ 106.0kPa^

^Note: required altitude for normal operation:

-400m ~ +3000m

Efficiency

Parameter

Linear range

Precision

Cross contamination

WBC

0-500×109 /L

≤2.5% (≥4.51×109 /L)

≤ 1.0%

RBC

0-8.60×1012/L

≤1.5% (≥3.5×1012/L)

≤ 1.0%

HGB

0-260g/L

≤1.0% (110-180g/L)

≤ 1.0%

HCT

0-75%

≤1.5% (30%-50%)

≤ 1.0%

 

PLT*

 

0-5000×109 /L

≤ 1.5(SD) (≤20×109/L)^

≤ 2.5% (≥100×109/L)^

 

≤ 1.0%

 

RET*

 

0-0.8×1012/L

≤15% (RBC ≥ 3.00×1012/L RET%: 1.00% ~ 4.00%)

 

≤ 1.0%

ESR

 

≤1.8(SD)(0~20mm/h)

≤ 1.0%

^ Note: Only applicable to CDR/PLT-O 5x and CR/PLT-O 5x series