Technical Features
System Functions:
-
Capacity: 600 photometric tests/hour, up to 800 tests/hour with ISE.
-
Optical Measurement Principle: Absorb, trim along the strip.
-
Number of Tests Per Time: Up to 77 tests.
-
Run Multiple Tests: 3 electrolyte tests + 3 photometric tests in just 5 bar cycles.
-
Measurement Method: Endpoint, fixed time, kinetics, ISE.
Model Management:
-
SDM: 120 samples on 12 trays, continuous sample loading, priority given to urgent samples.
-
Sample Volume: 1.5 – 45µl, 0.1µl increments.
-
HbA1C sample: Perform HbA1C test.
-
Interface: Sample ID, barcode, patient code, test code, chemical code.
Chemical Management:
-
Chemical Tray: 80 positions with cooling system (2 – 8°C).
-
Chemical Volume: 10 – 200µl, error 0.5µl.
-
Chemical Divider: Distinguish between fluid levels, clots and air bubbles.
Integrated Barcode Reader:
-
Barcode: Supports Codabar, ITF, Code 128, Code 39, UPC/EAN and Code 93.
-
LIS System Connection.
Reaction Chamber System:
-
Reaction Volume: 10 – 300µl.
-
Temperature Accuracy: ±0.1°C, fluctuation 0.1°C.
-
Recognition Response: 124 isothermal cuvettes with 8-step automatic washing cycle.
Optical System:
-
Light Source: Tungsten-halogen lamp.
-
Photometer: Grating photometer.
-
Wavelength: 16 wavelengths (340nm, 380nm, 412nm, 450nm, 480nm, 505nm, 546nm, 570nm, 605nm, 630nm, 660nm, 700nm, 740nm, 800nm, 850nm, 880nm).
-
Absorption Range: 0 – 3.5Abs (difference 0.1 nm).
-
Calibration and Verification:
-
Calibration: Linear (endpoint, fixed point and kinematic), Logit-log 4P, Logit-log 5P, Spline, Exponential, Polynomial, Parabola.
-
Law of Standardization: Westgard multi-rule, Twin plot.
-
Operations Department:
-
Operating System: Touch screen (optional), Windows 10/64bit.
-
Interface: RS-232, Network Port, USB/parallel port.
Operating Conditions:
-
Input Power Source: 200 – 240V, 50/60Hz, 1700VA.
-
Temperature: 10 – 35°C.
-
Humidity: 35 – 85%.
-
Size: 1380mm(length)860mm(width)1200mm(height).
-
Weight: 350kg.

Đánh giá